DVM1110 Specifications

The following general and physical technical specifications are provided for reference when operating the DVM1100 .

If you need to clean the module, wipe it with a dry cloth.

General

  • Warm-up time: no specified warmup time

  • Bus type: USB, slave

  • Altitude: Up to 2,000 m

  • Working Voltage:

    • 250 V maximum between Input/earth ground

    • 250 V maximum between COM/earth ground

    • 500 mV maximum shunt/COM

  • Power requirement: +5 V DC, 155 mA typical

  • Input Impedance: 100 MΩ on primary input | 100 kΩ on shunt input*

    *This input has a range and impedance specially designed for measuring shunts more accurately.

  • Measurement Category: CAT II

  • Pollution Degree: 1

  • Ingress protection rating: IP67

  • Accuracy by Range:

    Accuracy by Range

    Type

    Range

    Accuracy**

    Temp Coefficient

    DC

    500 mV

    5 V

    250 V

    ± (0.1% + 0.025 mV)†

    ± (0.1% + 0.001 V)†

    ± (0.1% + 0.05 V)†

    0.007% / °C

    0.007% / °C

    0.007% / °C

    AC

    350 mV RMS

    3.5 V RMS

    175 V RMS

    ± (1% + 0.105 mV)††

    ± (1% + 0.00105 V)††

    ± (1% + 0.0525 V)††

    N/A

    N/A

    N/A


  • ** Accuracy shown is for operation in temperatures from 15°C to 35°C. Annual calibration is required to maintain DVM accuracy.

    † DC voltage error increases by 0.1% in the presence of 35 to 100 volts of AC interference.

    †† AC voltage accuracy applies to input frequencies of 50 or 60 Hz.

  • Accuracy Examples:

    Accuracy Examples

    Type

    Signal

    Accuracy

    DC

    100 mV

    850 mV

    50 V

    ± 0.13 mV

    ± 1.9 mV

    ± 0.1 V

    AC

    100 mV

    850 mV

    50 V

    ± 1.1 mV

    ± 9.5 mV

    ± 0.55 V

Physical

  • Size and Weight:

    Physical Dimensions

    Type

    W (in.)

    L (in.)

    H (in.)

    Weight

    DVM2130

    2.94

    4.44

    1.50

    134g (4.8oz)

    DVM1110

    2.88

    4.00

    1.31

     

  • Operating temperature: 0°C to 55°C

  • Storage temperature: -40°C to 70°C

  • Relative humidity: 10% to 90% non-condensing