The following general and physical technical specifications are provided for reference when operating the DVM1110.
If you need to clean the module, wipe it with a dry cloth.
General
-
Warm-up time: TBD minutes for measurements accurate within specifications
-
Bus type: USB, slave
-
Altitude: Up to 2,000 m
-
Working Voltage:
-
250 V maximum between Input/earth ground
-
250 V maximum between COM/earth ground
-
500 mV maximum shunt/COM
-
-
Power requirement: +5 V DC, 155 mA typical
-
Input Impedance: 100 MΩ on primary input | 100 kΩ on shunt input*
*This input has a range and impedance specially designed for measuring shunts more accurately.
-
Measurement Category: CAT II
-
Pollution Degree: 1
-
Ingress protection rating: IP67
-
Accuracy by Range and Examples:
Accuracy by Range Accuracy by Range:
Accuracy Examples:
Range
Accuracy**
Temp Coefficient
Signal
Accuracy
DC
500 mV
5 V
250 V
± (0.1% + 0.025 mV)†
± (0.1% + 0.001 V)†
± (0.1% + 0.05 V)†
0.007% / °C
0.007% / °C
0.007% / °C
DC
100 mV
850 mV
50 V
± 0.13 mV
± 1.9 mV
± 0.1 V
AC
350 mV RMS
3.5 V RMS
175 V RMS
± (1% + 0.105 mV)††
± (1% + 0.00105 V)††
± (1% + 0.0525 V)††
N/A
N/A
N/A
AC
100 mV
850 mV
50 V
± 1.1 mV
± 9.5 mV
± 0.55 V
** Accuracy shown is for operation in temperatures from 15°C to 35°C. Annual calibration is required to maintain DVM accuracy.
† DC voltage error increases by 0.1% in the presence of 35 to 100 volts of AC interference.
†† AC voltage accuracy applies to input frequencies of 50 or 60 Hz.
Physical
-
Dimensions: 10.0 x 7.3 x 3.3 cm
-
Weight: 134 g (4.8 oz)
-
Operating temperature: 0°C to 55°C
-
Storage temperature: -40°C to 70°C
-
Relative humidity: 10% to 90% non-condensing