DVM1110 and DVM2130 Specifications
The DVM1110 (built into the Allegro AX) and the DVM2130 (a stand-alone DVM) include the ability to take readings as well as GPS location information.
The DVM1110 and DVM2130 include the following specifications.
General
Accuracy by Range |
||||||
---|---|---|---|---|---|---|
Range | Accuracy** | Temp Coefficient | Signal |
Accuracy |
||
DC | 500 mV | ± (0.1% + 0.025 mV)† | 0.007% / °C | DC | 100 Mv | ± 0.13 mV |
5 V | ± (0.1% + 0.001 V)† | 0.007% / °C | 850 mV | ± 1.9 mV | ||
250 V | ± (0.1% + 0.001 V)† | 0.007% / °C | 50 V | ±0.1 V | ||
AC | 350 mV | ± (0.1% + 0.05 V)† | N/A | AC | 100 mV | ± 1.1 mV |
3.5 V | ± (1% + 0.00105 V)†† | N/A | 850 mV | ± 9.5 mV | ||
175 V | ± (1% + 0.0525 V)†† | N/A | 50 V | ± 0.55 V | ||
*This input has a range and impedance specially designed for measuring shunts more accurately.
**Accuracy shown is for operation in temperatures from 15°C to 35°C. Annual calibration is required to maintain DVM accuracy.
†DC voltage error increases by 0.1% in the presence of 35 to 100 volts of AC interference.
††AC voltage accuracy applies to input frequencies of 50 or 60 Hz.
Physical
-
Size and Weight:
Physical Dimensions Type
W (in.)
L (in.)
H (in.)
DVM2130 2.94
4.44
1.50
DVM1110 2.88
4.00
1.31
-
Operating temperature: 0°C to 55°C
-
Storage temperature: -40°C to 70°C
-
Relative humidity: 10% to 90% non-condensing