DVM1110 and DVM2130 Specifications

The DVM1110 (built into the Allegro AX) and the DVM2130 (a stand-alone DVM) include the ability to take readings as well as GPS location information.

The DVM1110 and DVM2130 include the following specifications.

General

DVM1110/DVM2130 Specifications

Accuracy by Range

           
  Range Accuracy** Temp Coefficient   Signal

Accuracy

DC 500 mV ± (0.1% + 0.025 mV)† 0.007% / °C DC 100 Mv ± 0.13 mV
  5 V ± (0.1% + 0.001 V)† 0.007% / °C   850 mV ± 1.9 mV
  250 V ± (0.1% + 0.001 V)† 0.007% / °C   50 V ±0.1 V
AC 350 mV ± (0.1% + 0.05 V)† N/A AC 100 mV ± 1.1 mV
  3.5 V ± (1% + 0.00105 V)†† N/A   850 mV ± 9.5 mV
  175 V ± (1% + 0.0525 V)†† N/A   50 V ± 055 V
             

*This input has a range and impedance specially designed for measuring shunts more accurately.

**Accuracy shown is for operation in temperatures from 15°C to 35°C. Annual calibration is required to maintain DVM accuracy.

†DC voltage error increases by 0.1% in the presence of 35 to 100 volts of AC interference.

††AC voltage accuracy applies to input frequencies of 50 or 60 Hz.

Physical

  • Size and Weight:

    Physical Dimensions

    Type

    W (in.)

    L (in.)

    H (in.)

    DVM2130

    2.94

    4.44

    1.50

    DVM1110

    2.88

    4.00

    1.31

  • Operating temperature: 0°C to 55°C

  • Storage temperature: -40°C to 70°C

  • Relative humidity: 10% to 90% non-condensing